<PRE>
|
||||
$Id: m_dmm.C,v 1.8 2006/01/09 16:04:14 gef1sa Exp $
|
||||
---------------------------------------------------------------------
|
||||
Kurzbeschreibung: Aufruf des Prozesses VXI_DMM
|
||||
---------------------------------------------------------------------
|
||||
Projekt: CCS
|
||||
Sourcefile: m_dmm.C
|
||||
Autor: RtW/TEF72-Juilfs (12.12.95)
|
||||
Zielsystem: HP9000/V743 (HP-RT 2.0x), /Sxxx (HP-UX 9.0x)
|
||||
Sprache: C++
|
||||
---------------------------------------------------------------------
|
||||
Prototyp:
|
||||
MPI Schnittstelle
|
||||
void m_dmm(shm_stringpointer *zeilen_info
|
||||
,shm_stringpointer *zeilen_wert
|
||||
,long index1
|
||||
,long anzahl)
|
||||
|
||||
---------------------------------------------------------------------
|
||||
Aufruf: im Vectorinterpreter
|
||||
|
||||
(*funktionen[modulnummer])
|
||||
(Ssp_expand_zeile
|
||||
,Ssp_expand_wert
|
||||
,p_verwaltung->begin_modulblock
|
||||
,p_verwaltung->anzahl_modulparameter);
|
||||
|
||||
|
||||
|
||||
|
||||
---------------------------------------------------------------------
|
||||
Funktionsbeschreibung:
|
||||
Dieses Modul <20>bergibt seine Parameter an den Prozess VXI_DMM.
|
||||
Im Prozess VXI_DMM werden dann die Messungen durchgef<65>hrt.
|
||||
Das Me<4D>ergebnis wird per Returnparameter aus diesem Modul weiter
|
||||
an das Modul M_AUSWERTEN gegeben. Die Funktionalit<69>t des Multimeters
|
||||
wird mit den <20>bergabeparametern eingestellt.
|
||||
|
||||
|
||||
Beschreibung der Modul<75>bergabeparameter:</pre><b><a name="param1">1</a> PRUEFSCHRITT</b><pre>
|
||||
GAUDI Pr<50>fschritt
|
||||
------------------------------</pre><b><a name="param2">2</a> POSITION</b><pre>
|
||||
PAV- Position
|
||||
------------------------------</pre><b><a name="param3">3</a> VXI_GERAET</b><pre> Name des Ger<65>tes laut PAM Beschreibungsfile
|
||||
:DVM_01
|
||||
:DVM_02
|
||||
------------------------------</pre><b><a name="param4">4</a> DMM_FUNKTION</b><pre> Me<4D>funktion
|
||||
:DCV
|
||||
:ACV
|
||||
:RESISTANCE
|
||||
------------------------------</pre><b><a name="param5">5</a> DMM_RANGE</b><pre>
|
||||
Messbereich in Volt Kohm
|
||||
0.0 Autorange (Sonderfunktion)
|
||||
|
||||
------------------------------</pre><b><a name="param6">6</a> DMM_APER_TIME</b><pre> Integrationszeit in s
|
||||
Folgende Integrationszeiten k<>nnen eingestellt werden.
|
||||
10 us, 100us, 2,5 ms, 266, ms
|
||||
sollte ein Integrationswerteingabe zwischen 2 Werten liegen, wird immer der n<>here
|
||||
verwendet.
|
||||
|
||||
|
||||
|
||||
------------------------------</pre><b><a name="param7">7</a> DMM_TRG_SOURCE</b><pre> Triggerart
|
||||
:IMMEDIATE interner Trigger
|
||||
:EXT externer Trigger
|
||||
:HOLD Halten des Me<4D>ergebnisses
|
||||
------------------------------</pre><b><a name="param8">8</a> DMM_TRG_COUNT</b><pre> Anzahl der Triggerimpulse
|
||||
------------------------------</pre><b><a name="param9">9</a> DMM_TRG_DELAY</b><pre> Delay zwischen 1. Trigger und 1.Messung
|
||||
------------------------------</pre><b><a name="param10">10</a> DMM_SMP_SOURCE</b><pre> Quelle des Triggerimpulses
|
||||
:IMMEDIATE Interner Trigger
|
||||
:TIMER Keine Beschreibung
|
||||
:SOFTWARE
|
||||
:CONTROLREG
|
||||
------------------------------</pre><b><a name="param11">11</a> DMM_SMP_COUNT</b><pre> Anzahl von Messungen per Trigger
|
||||
-------------------------------</pre><b><a name="param12">12</a> DMM_SMP_PERIOD</b><pre> Zeitdauer zwischen 2 Messungen
|
||||
-------------------------------</pre><b><a name="param13">13</a> DMM_AUTOZERO</b><pre> :ON
|
||||
:OFF
|
||||
:ONCE
|
||||
-------------------------------</pre><b><a name="param14">14</a> DMM_OFFSET_COM</b><pre> :TRUE
|
||||
:FALSE
|
||||
-------------------------------</pre><b><a name="param15">15</a> DMM_INTERRUPT</b><pre> -------------------------------</pre><b><a name="param16">16</a> DMM_MASKE</b><pre> -------------------------------</pre><b><a name="param17">17</a> DMM_MODUS</b><pre> Steuerfunktion des DMM Einstellen Arm Lesen
|
||||
INIT +
|
||||
ARM +
|
||||
READ +
|
||||
S + +
|
||||
AR (f<>r ACV) + +
|
||||
M (= NORM) + + +
|
||||
|
||||
-------------------------------</pre><b><a name="param18">18</a> R_MESSWERT</b><pre> R<>ckgabe des Messwertes
|
||||
-------------------------------</pre><b><a name="param19">19</a> TIMEOUT</b><pre></pre><b><a name="param20">20</a> DIMENSION</b><pre></pre><b><a name="param21">21</a> MESSTEILERFAKTOR</b><pre>
|
||||
|
||||
|
||||
|
||||
######################################################################
|
||||
Status:
|
||||
Bisher wurden beim Aufruf von ACVINIT 2 Sekunden gewartet. Dies
|
||||
soll mit einer neuen Prozessstruktur ge<67>ndert werden. Die Funktion
|
||||
RACV war dann relativ schnell <200 ms.
|
||||
Die Pr<50>fzeit erh<72>hte sich dadurch um 2 Sekunden
|
||||
|
||||
<20>nderung des Ablauf:
|
||||
Ab der CCS Release 5.2 wird ist es m<>glich f<>r das 2. DVM einen eigenen Prozess
|
||||
aufzusetzen. Hiermit sollen die ACVINIT Aufrufe in den Backround Prozess
|
||||
abgearbeitet werden.
|
||||
|
||||
|
||||
|
||||
Hieraus ergeben sich folgende F<>lle.
|
||||
1.) Es gibt nur einen Prozess VD
|
||||
Bei Aufruf des ACVINIT wird der VI sofort wieder freigegeben und beim n<>chsten
|
||||
Aufruf von M_DMM die restliche Initialisierunszeitzeit gewartet.
|
||||
Dies bedeutet, dass ggf. beim n<>chsten DVM Aufruf DCV 2 Sekunden gewartet wird.
|
||||
Grund: Die ACVINIT ist noch nicht abgeschlossen.
|
||||
|
||||
2.) Es gibt einen weiteren Prozess VD ---> V2
|
||||
Bei Aufruf des ACVINIT wird der VI sofort wieder freigegeben und beim n<>chsten
|
||||
Aufruf von M_DMM die restliche Initialisierunszeitzeit gewartet.
|
||||
Da die DCV Messungen klassisch mit dem DVM_01 durchgef<65>hrt werden,
|
||||
und das DVM_01 <20>ber dem Prozess VD gesteuert wird, gibt es keinen
|
||||
Performance Verlust. Beim Aufsetzen des RACV wird eventuell die Initialisierung
|
||||
Zeit des ACVINIT noch abgewartet.
|
||||
Der Pr<50>flisten Autor mu<6D> darauf achten das zwischen Aufruf ACVINIT und RACV
|
||||
ca 2 Sekunden in der Pr<50>fung verbraucht werden.
|
||||
bei diesem Verfahren geht die InitialisierungsZeit nicht in die Pr<50>fzeit ein,
|
||||
da der Prozess V2 eigenst<73>ndig das DVM_02 initialisiert, und vor der Initialisierung
|
||||
den VI freigegeben hat.
|
||||
Beim Umschalten des V2 Prozesses auf DCV Messungen mu<6D> nat<61>rlich wieder mit
|
||||
einer uminitialisierungszeit von ca 300 ms gerechnet werden. Dies war aber
|
||||
auch schon vorher der Fall.
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
</PRE>
|
||||